Simple and rapid determination of PCDD/Fs in flue gases from various waste incinerators in China using DR-EcoScreen
Zhiguang Zhou1, Bin Zhao2, Hiroyuki Kojima3
1State Key Laboratory of Environmental Chemistry and Ecotoxicology, Research Center for Eco-Environmental Sciences, Chinese Academy of Sciences, Beijing 100085, China; State Environmental Protection Key Laboratory of Dioxin Pollution Control, National Research Center for Environmental Analysis and Measurement, 1 South Yuhui Rd, Chaoyang District, Beijing 100029, China.
Abstract:
In developing countries such as China, there is a strong need for simple and rapid bioassays for the determination of polychlorinated dibenzo-p-dioxins and dibenzofurans (PCDD/Fs) in environmental samples; i.e., flue gas and fly ash from waste incinerators. In this study, we applied the DR-EcoScreen cell (DR-cell) assay to determination of PCDD/Fs in 78 flue gas samples obtained from various waste incinerators in China between 2009 and 2011. The flue gas samples were obtained from four kinds of incinerators, classified into hazardous, medical and municipal-solid waste, and iron ore sintering, and the flue gas extracts were cleaned up using an SPD-600 automated-sample preparation device for DR-cell assay. The PCDD/Fs values obtained from the DR-cell assay were compared with those obtained from conventional high resolution gas chromatography-high resolution mass spectrometry (HRGC-HRMS) analysis. The bioanalytical equivalent (BEQ) values obtained from the DR-cell assay were very closely correlated with the international toxicity equivalent (I-TEQ) values from HRGC-HRMS analysis (r2=0.98, n=78), while the BEQ values were 5.52-fold higher than the I-TEQ values, as the PCDFs, which account for 80% of the total I-TEQ value, were overestimated by DR cell-assay. Therefore, we multiplied the BEQ values from the DR-cell assay by a conversion coefficient (0.181, the reciprocal of 5.52), and could approximate the TEQ values from the HRGC-HRMS analysis. These results suggest that the DR-cell assay combined with SPD-600 cleanup provides a promising method for the simple and rapid screening of PCDD/Fs levels in flue gas samples, such as those from various waste incinerators in China.
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...