Insulation failure with externalized conductor of a Linox SD lead: a case report
Joseph A Manfredi1, Steven M Smithgall1, Charles M Kircher1
1AnMed Health Arrhythmia Specialists, Anderson, South Carolina, USA.
Abstract:
A 53-year-old man with a dual-chamber implantable cardioverter-defibrillator presented to the electrophysiology (EP) lab for an EP study in evaluation of supraventricular tachycardia despite medical therapy. Fluoroscopic imaging during placement of diagnostic catheters demonstrated insulation failure with an externalized conductor on the right ventricular defibrillator lead just distal to the proximal defibrillator coil. Device interrogation revealed normal pacing, and shock impedances, normal sensing, and stable pacing thresholds.
More Related Videos
Related Concept Videos
Insulation Coordination
Charging Conductors By Induction
Generally, conductors like metals do not allow any excess charge to be present on them. Any excess charge added to metals easily flows away, for example, when a metal is placed on the Earth. This process is called earthing.
However, conductors can be charged by a process called induction. For example, consider charging a...
Charge on a Conductor
Electric Field Inside a Conductor
Suppose a piece of metal is placed near a positive charge. The free electrons in the metal are attracted to the external positive charge and migrate freely toward that region. This region then...
Electrostatic Boundary Conditions in Dielectrics
Consider a case where both the mediums across a boundary are two different dielectric materials. Recall that the electric field and electric displacement are proportional and related through the material's permittivity....
Resistance and Conductance
Various factors impact the resistance of a conductor. Spiraling in stranded conductors increases their...