Related Experiment Video
Updated: May 4, 2026

Preoxygenation Techniques for Tracheal Intubation in Critically Ill Adults Utilizing Oxygen Mask and Noninvasive Ventilation
Published on: December 5, 2025
Comparative evaluation of three interfaces for non-invasive ventilation: a randomized cross-over design physiologic
Introduction:
Interface choice is crucial for non-invasive ventilation (NIV) success. We compared a new interface, the helmet next (HN), with the facial mask (FM) and the standard helmet (HS) in twelve healthy volunteers.
Methods:
In this study, five NIV trials were randomly applied, preceded and followed by a trial of unassisted spontaneous breathing (SB). Baseline settings, for example, 5 cmH2O of both inspiratory pressure support (PS) and positive end-expiratory pressure (PEEP), were applied through FM, HS and HN, while increased settings (PS and PEEP of 8 cmH2O) were only applied through HS and HN. We measured flow, airway, esophageal and gastric pressures, and calculated inspiratory effort indexes and trigger delays. Comfort was assessed with a visual-analog-scale.
Results:
We found that FM, HS and HN at baseline settings were not significantly different with respect to inspiratory effort indexes and comfort. Inspiratory trigger delay and time of synchrony (TI,synchrony) were significantly improved by FM compared to both helmets, whereas expiratory trigger delay was shorter with FM, as opposed to HS only. HN at increased settings performed better than FM in decreasing inspiratory effort measured by pressure-time product of transdiaphragmatic pressure (PTPdi)/breath (10.7 ± 9.9 versus 17.0 ± 11.0 cmH2Os), and PTPdi/min (128 ± 96 versus 204 ± 81 cmH2Os/min), and PTPdi/L (12.6 ± 9.9 versus 30.2 ± 16.8 cmH2Os/L). TI, synchrony was inferior between HN and HS at increased settings and FM.
Conclusions:
HN might hold some advantages with respect to interaction and synchrony between subject and ventilator, but studies on patients are needed to confirm these findings.
Trial Registration:
ClinicalTrials.gov NCT01610960.
More Related Videos
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Bioavailability Study Design: Healthy Subjects Versus Patients