Related Experiment Video
Updated: May 4, 2026

Evaluation of Drug Sorption to PVC- and Non-PVC-based Tubes in Administration Sets Using a Pump
Published on: March 11, 2017
Impact of infusion set characteristics on the accuracy of patient-controlled morphine administration: a controlled
A Foinard1, B Décaudin, C Barthélémy
1Department of Biopharmacy, Galenic and Hospital Pharmacy, Lille 2 University, UDSL, EA GRIIOT, UFR Pharmacie, Lille, France.
Abstract:
The aim of our in-vitro study was to assess the impact of infusion set characteristics on the accuracy of morphine doses in patient-controlled analgesia. Two infusion sets differing in conception and dead-space volume were assessed: a standard set and a low dead-space volume Y-set. The patient-controlled analgesia programme parameters were as follows: bolus equal to 1 ml at 100 ml.h(-1) ; lockout intervals equal to 5 and 10 min; and carrier fluid flow rate equal to 10 and 50 ml.h(-1) . Morphine concentration was determined by an ultraviolet spectrophotometric method. The morphine doses were significantly different from one set to the other during bolus and lockout intervals, whatever the patient-controlled analgesia programme. The average doses were approximately 1.3-6.0 times higher with the low dead-space volume Y-set during bolus. Our study underlines the impact of infusion set characteristics on the accuracy of morphine patient-controlled analgesia doses.
Related Concept Videos
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
One-Compartment Open Model for IV Bolus Administration: General Considerations
The drug's presence in the body is defined by an equation representing the difference between the rates of drug entry and exit. Key parameters—elimination rate constant,...
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions
IV Infusion to Oral Dosing: Conversion Methods
Methods for Studying Drug Absorption: In situ
The Doluisio method involves perfusing a prepared segment of a rat's small intestine with a solution of radiolabeled drug and a non-absorbable marker. This helps to differentiate between absorbed and non-absorbed drug concentrations. The intestinal segment is connected at both ends using tubing and syringes,...