Related Experiment Video
Updated: May 4, 2026

A High-Throughput Electrochemiluminescence 7-Plex Assay Simultaneously Screening for Type 1 Diabetes and Multiple Autoimmune Diseases
Published on: May 29, 2020
Causes of false-positive HIV rapid diagnostic test results
Abstract:
HIV rapid diagnostic tests have enabled widespread implementation of HIV programs in resource-limited settings. If the tests used in the diagnostic algorithm are susceptible to the same cause for false positivity, a false-positive diagnosis may result in devastating consequences. In resource-limited settings, the lack of routine confirmatory testing, compounded by incorrect interpretation of weak positive test lines and use of tie-breaker algorithms, can leave a false-positive diagnosis undetected. We propose that heightened CD5+ and early B-lymphocyte response polyclonal cross-reactivity are a major cause of HIV false positivity in certain settings; thus, test performance may vary significantly in different geographical areas and populations. There is an urgent need for policy makers to recognize that HIV rapid diagnostic tests are screening tests and mandate confirmatory testing before reporting an HIV-positive result. In addition, weak positive results should not be recognized as valid except in the screening of blood donors.
More Related Videos
07:40Field Postmortem Rabies Rapid Immunochromatographic Diagnostic Test for Resource-Limited Settings with Further Molecular Applications
Published on: June 29, 2020
08:37Development of a Quantitative Recombinase Polymerase Amplification Assay with an Internal Positive Control
Published on: March 30, 2015