Related Experiment Video
Updated: May 4, 2026

Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
CT angiography-fluoroscopy fusion imaging for percutaneous transapical access
Chad Kliger1, Vladimir Jelnin1, Sonnit Sharma1
1Department of Cardiovascular Medicine, Division of Structural and Congenital Heart Disease, Lenox Hill Heart and Vascular Institute-North Shore/Long Island Jewish Health System, New York, New York.
Objectives:
The aim of this proof-of-principle study is to validate the accuracy of fusion imaging for percutaneous transapical access (TA).
Background:
Structural heart disease interventions, including TA, are commonly obtained under fluoroscopic guidance, which lacks important spatial information. Computed tomographic angiography (CTA)-fluoroscopy fusion imaging can provide the 3-dimensional information necessary for improved accuracy in planning and guidance of these interventions.
Methods:
Twenty consecutive patients scheduled for percutaneous left ventricular puncture and device closure using CTA-fluoroscopy fusion guidance were prospectively recruited. The HeartNavigator software (Philips Healthcare, Best, the Netherlands) was used to landmark the left ventricular epicardium for TA (planned puncture site [PPS]). The PPS landmark was compared with the position of the TA closure device on post-procedure CTA (actual puncture site). The distance between the PPS and actual puncture site was calculated from 2 fixed reference points (left main ostium and mitral prosthesis center) in 3 planes (x, y, and z). The distance from the left anterior descending artery at the same z-plane was also assessed. TA-related complications associated with fusion imaging were recorded.
Results:
The median (interquartile range [IQR]) TA distance difference between the PPS and actual puncture site from the referenced left main ostium and mitral prosthesis center was 5.00 mm (IQR: 1.98 to 12.64 mm) and 3.27 mm (IQR: 1.88 to 11.24 mm) in the x-plane, 4.48 mm (IQR: 1.98 to 13.08 mm) and 4.00 mm (IQR: 1.62 to 11.86 mm) in the y-plane, and 5.57 mm (IQR: 3.89 to 13.62 mm) and 4.96 mm (IQR: 1.92 to 11.76 mm) in the z-plane. The mean TA distance to the left anterior descending artery was 15.5 ± 7.8 mm and 22.7 ± 13.7 mm in the x- and y-planes. No TA-related complications were identified, including evidence of coronary artery laceration.
Conclusions:
With the use of CTA-fluoroscopy fusion imaging to guide TA, the actual puncture site can be approximated near the PPS. Moreover, fusion imaging can help maintain an adequate access distance from the left anterior descending artery, thereby, potentially reducing TA-related complications.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization I: Pre-Procedure Overview
Imaging Studies VII: Vascular Imaging