Systematic evaluation of evidence on veterinary viscoelastic testing part 1: System comparability
Maureen McMichael1, Robert Goggs, Stephanie Smith
1Department of Veterinary Clinical Medicine, College of Veterinary Medicine.
Results from thromboelastography and rotational thromboelastometry instruments are not directly comparable. Avoid extrapolating data between machines and standardize preanalytical variables for reliable point-of-care testing.
Area of Science:
- Veterinary Medicine
- Point-of-Care Diagnostics
- Hemostasis Monitoring
Background:
- Thromboelastography (TEG) and rotational thromboelastometry (ROTEM) are viscoelastic point-of-care instruments used for hemostasis monitoring.
- Assessing the comparability of results between these two systems is crucial for accurate clinical decision-making.
Purpose of the Study:
- To systematically review the existing evidence regarding the comparability of TEG and ROTEM systems.
- To identify any knowledge gaps concerning the interchangeability of data generated by these analyzers.
Main Methods:
- A systematic literature evaluation was conducted using databases such as Medline, PubMed, CAB abstracts, and Google Scholar.
- Relevant articles were categorized by their level of evidence and quality.
- Consensus was developed for applying findings to clinical practice.
Main Results:
- Eight relevant articles were identified and analyzed.
- None of the studies supported the direct comparability of the two systems.
- One study was neutral (Level of Evidence 6, Poor quality), and seven studies opposed comparability (varying levels of evidence and quality).
Conclusions:
- Results generated by TEG and ROTEM analyzers are not directly comparable.
- Extrapolation of results between these machines should be avoided.
- Standardization of preanalytical variables (e.g., blood collection, holding time, temperature) is strongly recommended.
- Each institution should establish site-specific reference values for each machine.