Three-Dimensional Imaging of Aortic Tissues in Atherosclerosis
Published on: October 25, 2024
Characterization of atherosclerotic plaque: a contrast-detail study using multidetector and cone-beam computed
Nima Kasraie1, Peter Mah, Carl R Keener
1The University of Texas Health Science Center at San Antonio. Kasraie@livemail.uthscsa.edu.
Abstract:
A Hindmarsh-Rose model perceptibility phantom containing inserts with various in vitro atherosclerotic plaque compositions was constructed and imaged on a clinical 64 slice multidetector (MDCT) system using 80 and 120 kVp settings and two other cone-beam (CBCT) systems at 80 kVp. Perceptibility of the simulated lipid-fibrotic plaque solutions in the images was evaluated by six observers. The effective doses of the protocols employed were estimated using phantom CTDI-vol measurements placed at identical settings. The CBCT system allowed reduction in effective dose in comparison with the conventional MDCT system for imaging of the carotid plaque phantoms without degrading image quality. The CBCT dose was less than MDCT, with a mean dose of 1.14 ± 0.01 mSv and 1.11 ± 0.02 mSv for MDCT using two measuring techniques vs. 0.35 ± 0.01 mSv for CBCT. The image quality analysis showed no significant differences in the contrast-detail (C-D) curves of the best performing CBCT vs. clinical MDCT (p > 0.05) using a Mann-Whitney U test. Results indicate that low-tube-potential CBCT may produce comparable C-D resolution for phantom-based representations of soft plaque types with respect to MDCT systems. This study suggests that the utility of low kVp CT techniques for evaluating carotid vulnerable atherosclerotic plaque merits further study.
More Related Videos
13:45A Method to Study the Correlation Between Local Collagen Structure and Mechanical Properties of Atherosclerotic Plaque Fibrous Tissue
Published on: November 11, 2022
11:45Generation and 3-Dimensional Quantitation of Arterial Lesions in Mice Using Optical Projection Tomography
Published on: May 26, 2015
Related Concept Videos
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System V: CT
Atherosclerosis I: Introduction
Imaging Studies III: Computed Tomography
Atherosclerosis II: Clinical Manifestations and Diagnostic Tests
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...