Related Experiment Video
Updated: May 4, 2026

An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Interfacial sliding properties of bone screw materials and their effect on screw fixation strength
Arto P Koistinen1, Hannu Korhonen, Heikki Kröger
1Department of Applied Physics, University of Eastern Finland, Kuopio - Finland and SIB Labs, University of Eastern Finland, Kuopio - Finland.
Background:
This study examined the effect of interfacial sliding and test material properties on the fixation strength and insertional properties of self-tapping bone screws. Various substitute materials (polyacetal [POM], poly(methyl methacrylate) [PMMA] and E-glass-filled Epoxy [Sawbones®]) for human bone were evaluated, and the results were compared with the findings for cadaver bone.
Methods:
Initial coefficient of friction (CoF) of the screw material stainless steel AISI316 was tested using a pin-on-disk apparatus, and the screws were exposed to pullout tests after insertion torque tests. The effect of a smooth diamond-like carbon (DLC) coating was studied by applying the coating on both CoF test balls and bone screws.
Results:
Mechanical properties of test blocks strongly correlated to both pullout strength and insertion torque of the screws: for noncoated 2.7-mm screws, tensile strength correlated to pullout strength and insertion torque, with Pearson correlation coefficients r=0.977 and r=0.738, respectively. In contrast, CoF correlated strongly to screw insertion torque but not to pullout strength in bone substitute materials (for noncoated 2.7-mm screws, r=0.652 and r=0.248, respectively). There were no significant differences in CoF using noncoated and DLC-coated screw materials against bone substitutes.
Conclusions:
Proper materials for in vitro testing help in evaluating the biomechanics of the implants in advance. However, choosing the material needs attention, as their ability to model human bone depends on test type.
More Related Videos
09:07An Anesthesia, Surgery, and Harvest Method for the Evaluation of Transpedicular Screws Using an In Vivo Porcine Lumbar Spine Model
Published on: May 31, 2017
06:41A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
Related Concept Videos
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Screw: Problem Solving