Related Experiment Video
Updated: May 4, 2026

An Automated System for Sound Localization Testing in Hearing-Impaired Listeners
Published on: March 13, 2026
Biological calibration for web-based hearing tests: evaluation of the methods.
Marcin Masalski1, Tomasz Grysiński, Tomasz Kręcicki
1Department and Clinic of Otolaryngology, Head and Neck Surgery, Wroclaw Medical University, Wrocław, Poland. marcin.masalski@pwr.wroc.pl.
Biological calibration for online hearing tests using personal computers (PCs) is essential. The Bekesy method, with independent reference sound level determination per frequency, offers the lowest error for home-based hearing assessments.
Area of Science:
- Audiology
- Biomedical Engineering
- Digital Health
Background:
- Online hearing tests require accurate sound level calibration for reliable results.
- Biological calibration uses the hearing threshold of individuals with normal hearing to approximate reference sound levels.
- Personal computer (PC) based hearing tests conducted at home necessitate robust calibration methods.
Purpose of the Study:
- To evaluate the error, duration, and subjective difficulty of proposed biological calibration methods for PC-based online hearing tests.
- To identify optimal calibration techniques for home-based audiological assessments.
- To compare different methods for measuring calibration coefficients and determining reference sound levels.
Main Methods:
- Seven methods for measuring calibration coefficients were assessed, referencing normal hearing thresholds.
- Three methods were used to determine reference sound levels based on calibration coefficients.
- Web-based self-assessment of calibration coefficients was conducted in clinic and home settings, with pure-tone audiometry for comparison.
Main Results:
- The Bekesy audiometry method with a modulated signal showed the smallest calibration coefficient standard deviation (3.87 dB).
- Self-adjustment to the barely audible level was the simplest and shortest method, with a deviation of 4.97 dB.
- Methods determining reference sound levels independently for each frequency yielded the smallest error, with standard deviations ranging from 7.27 dB to 10.38 dB compared to pure-tone audiometry.
Conclusions:
- The study analyzed calibration error, time, and difficulty for web-based hearing tests.
- Results guide the selection of calibration methods for home-based audiological assessments.
- The Bekesy method with independent frequency-specific reference sound level determination is recommended for minimal error when time is not a constraint.
More Related Videos
04:32Sound Source Localization Testing in Single-sided Deafness Following Bone Conduction Intervention
Published on: December 20, 2024
14:05Behavioral Assessment of Hearing in 2 to 4 Year-old Children: A Two-interval, Observer-based Procedure Using Conditioned Play-based Responses
Published on: January 23, 2017