Related Experiment Video
Updated: May 4, 2026

Novel Percutaneous Approach for Deployment of 3D Printed Coronary Stenosis Implants in Swine Models of Ischemic Heart Disease
Published on: February 18, 2020
Coronary sinus cannulation with a steerable catheter during biventricular device implantation
Lingwei Wang1, Shiwen Yuan, Rasmus Borgquist
1Department of Arrhythmias, Skane University Hospital, Lund University , Lund , Sweden.
Objectives:
To determine whether a steerable catheter with electrogram guidance (CS-assist group) could facilitate access to the coronary sinus (CS) during cardiac resynchronization therapy (CRT) implantation.
Design:
Consecutive patients who underwent CRT implantation were recruited prospectively into the CS-assist group (n = 81) and compared with those using conventional techniques without an electrogram guidance (conventional group, n = 101).
Results:
The CS cannulation success rate was clearly greater in the CS-assist group (100%) than that in the conventional group (95%, p < 0.05), with significantly shorter mean procedure time (52.6 ± 20.6 min vs. 73.2 ± 40.9 min, p < 0.01) and fluoroscopy time (3.6 ± 3.2 min vs. 14.2 ± 20.4 min, p < 0.01). In the five CS cannulation failure cases, mean procedure time (144.0 ± 37.0 min) and fluoroscopy time (57.8 ± 24.8 min) were significantly longer than those in the other patients (61.2 ± 32.3 and 8.2 ± 13.6 min, respectively, n = 177, both p < 0.01).
Conclusions:
Using the steerable catheter with real-time electrogram guidance, location of and access to the CS is more rapid and successful, which may improve the success of the CRT implantation and may give significant time savings.