Related Experiment Video
Updated: May 3, 2026

Laryngeal Mask Airway LMA Placement in a Neonatal Patient Simulator Using a Non-Inflatable Supraglottic Airway SGA
Published on: July 14, 2023
Basic life support trained nurses ventilate more efficiently with laryngeal mask supreme than with facemask or
Elisabeth Gruber1, Rosmarie Oberhammer1, Karla Balkenhol2
1Department of Anaesthesiology and Critical Care Medicine, Hospital of Bruneck, Spitalstrasse 11, 39031 Bruneck, Italy.
Objective:
In some emergency situations resuscitation and ventilation may have to be performed by basic life support trained personnel, especially in rural areas where arrival of advanced life support teams can be delayed. The use of advanced airway devices such as endotracheal intubation has been deemphasized for basically-trained personnel, but it is unclear whether supraglottic airway devices are advisable over traditional mask-ventilation.
Methods:
In this prospective, randomized clinical single-centre trial we compared airway management and ventilation performed by nurses using facemask, laryngeal mask Supreme (LMA-S) and laryngeal tube suction-disposable (LTS-D). Basic life support trained nurses (n=20) received one-hour practical training with each device. ASA 1-2 patients scheduled for elective surgery were included (n=150). After induction of anaesthesia and neuromuscular block nurses had two 90-second attempts to manage the airway and ventilate the patient with volume-controlled ventilation.
Results:
Ventilation failed in 34% of patients with facemask, 2% with LMA-S and 22% with LTS-D (P<0.001). In patients who could be ventilated successfully mean tidal volume was 240±210 ml with facemask, 470±120 ml with LMA-S and 470±140 ml with LTS-D (P<0.001). Leak pressure was lower with LMA-S (23.3±10.8 cm H2O, 95% CI 20.2-26.4) than with LTS-D (28.9±13.9 cm·H2O, 95% CI 24.4-33.4; P=0.047).
Conclusions:
After one hour of introductory training, nurses were able to use LMA-S more effectively than facemask and LTS-D. High ventilation failure rates with facemask and LTS-D may indicate that additional training is required to perform airway management adequately with these devices. High-level trials are needed to confirm these results in cardiac arrest patients.
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...