Influence of contrast media viscosity and temperature on injection pressure in computed tomographic angiography: a
Madeleine Kok1, Casper Mihl, Alma A Mingels
1From the *Department of Radiology, †CARIM School for Cardiovascular Diseases, Departments of ‡Clinical Chemistry, and §Cardiology, Maastricht University Medical Center, Maastricht, The Netherlands; Departments of ∥Neuroradiology, and ¶Nuclear Medicine, University Hospital RWTH Aachen, Aachen, Germany.
Purpose:
Iodinated contrast media (CM) in computed tomographic angiography is characterized by its concentration and, consecutively, by its viscosity. Viscosity itself is directly influenced by temperature, which will furthermore affect injection pressure. Therefore, the purposes of this study were to systematically evaluate the viscosity of different CM at different temperatures and to assess their impact on injection pressure in a circulation phantom.
Materials And Methods:
Initially, viscosity of different contrast media concentrations (240, 300, 370, and 400 mgI/mL) was measured at different temperatures (20°C-40°C) with a commercially available viscosimeter. In the next step, a circulation phantom with physical conditions was used. Contrast media were prepared at different temperatures (20°C, 30°C, 37°C) and injected through a standard 18-gauge needle. All other relevant parameters were kept constant (iodine delivery rate, 1.9 g I/s; total amount of iodine, 15 g I). Peak flow rate (in milliliter per second) and injection pressure (psi) were monitored. Differences in significance were tested using the Kruskal-Wallis test (Statistical Package for the Social Sciences).
Results:
Viscosities for iodinated CM of 240, 300, 370, and 400 mg I/mL at 20°C were 5.1, 9.1, 21.2, and 28.8 mPa.s, respectively, whereas, at 40°C, these were substantially lower (2.8, 4.4, 8.7, and 11.2 mPa.s). In the circulation phantom, mean (SD) peak pressures for CM of 240 mg I/mL at 20°C, 30°C, and 37°C were 107 (1.5), 95 (0.6), and 92 (2.1) psi; for CM of 300 mg I/mL, 119 (1.5), 104 (0.6), and 100 (3.6) psi; for CM of 370 mg I/mL, 150 (0.6), 133 (4.4), and 120 (3.5) psi; and for CM of 400 mg I/mL, 169 (1.0), 140 (2.1), and 135 (2.9) psi, respectively, with all P values less than 0.05.
Conclusions:
Low concentration, low viscosity, and high temperatures of CM are beneficial in terms of injection pressure. This should also be considered for individually tailored contrast protocols in daily routine scanning.
More Related Videos
09:57Development and Evaluation of 3D-Printed Cardiovascular Phantoms for Interventional Planning and Training
Published on: January 18, 2021
12:15Tissue Preparation Techniques for Contrast-Enhanced Micro Computed Tomography Imaging of Large Mammalian Cardiac Models with Chronic Disease
Published on: February 8, 2022
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies VII: Vascular Imaging
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...