Inconsistency between simultaneous blood pressure measurements in the arm, forearm, and leg in anesthetized children
Ilan Keidan1, Avner Sidi2, Erez Ben-Menachem3
1Department of Anesthesiology and Intensive Care, Sheba Medical Center, Tel Hashomer,Israel (affiliated with Tel Aviv University, Sackler Faculty of Medicine, Tel Aviv, Israel); Department of Anesthesiology, College of Medicine, University of Florida, Gainesville, FL 32610-0254, USA.
Study Objective:
To determine the accuracy and precision of simultaneous noninvasive blood pressure (NIBP) measurement in the arm, forearm, and ankle in anesthetized children.
Design:
Prospective, randomized study.
Setting:
University medical center.
Patients:
101 ASA physical status 1 and 2 children (aged 1-8 yrs) scheduled for elective surgery with general anesthesia.
Measurements:
Simultaneous NIBP measurements were recorded at the arm, forearm, and ankle at 5-minute intervals.
Main Results:
The systolic blood pressure difference between the arm-forearm or the arm-ankle was within the ± 10% range in 63% and 29% of measurements, and within the ± 20% range in 85% and 67% of measurements, respectively. The diastolic blood pressure difference between the arm-forearm or the arm-ankle was within the ± 10% range in 42% and 44% and within the ± 20% range in 67% and 74% of measurements, respectively. In patients in whom the initial three NIBP measurements were within the ± 20% range between the forearm and arm, 86% of the subsequent measurements were also within that limit.
Conclusions:
Forearm and ankle NIBP measurements are unreliable and inconsistent with NIBP measured in the arm of anesthetized children. These alternative BP measurement sites are not reliable in accuracy (comparison with reference "gold" standard) and precision (reproducibility).
More Related Videos
07:35Simultaneous Electrocardiography Recording and Invasive Blood Pressure Measurement in Rats
Published on: January 31, 2019
06:51Development of an Algorithm to Perform a Comprehensive Study of Autonomic Dysreflexia in Animals with High Spinal Cord Injury Using a Telemetry Device
Published on: July 29, 2016
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pre-Procedural Guidelines for Assessing Blood Pressure
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure in the Leg
Preparation: