Related Experiment Video
Updated: May 3, 2026

Precision Measurements and Parametric Models of Vertebral Endplates
Published on: September 17, 2019
Handheld real-time volumetric imaging of the spine: technology development
Mohamed Tiouririne1, Sarah Nguyen, John A Hossack
1Department of Anesthesiolgoy, Division of Obstetric Anesthesia , Univeristy of Virginia, Charlottesville, VA , USA .
Abstract:
Technical difficulties, poor image quality and reliance on pattern identifications represent some of the drawbacks of two-dimensional ultrasound imaging of spinal bone anatomy. To overcome these limitations, this study sought to develop real-time volumetric imaging of the spine using a portable handheld device. The device measured 19.2 cm × 9.2 cm × 9.0 cm and imaged at 5 MHz centre frequency. 2D imaging under conventional ultrasound and volumetric (3D) imaging in real time was achieved and verified by inspection using a custom spine phantom. Further device performance was assessed and revealed a 75-min battery life and an average frame rate of 17.7 Hz in volumetric imaging mode. The results suggest that real-time volumetric imaging of the spine is a feasible technique for more intuitive visualization of the spine. These results may have important ramifications for a large array of neuraxial procedures.