Integrated circuit detector technology in abdominal CT: added value in obese patients
Fabian Morsbach1, Sebastian Bickelhaupt, Susan Rätzer
11 Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, Raemistrasse 100, CH-8091 Zurich, Switzerland.
Objective:
The purpose of this article was to assess the effect of an integrated circuit (IC) detector for abdominal CT on image quality.
Materials And Methods:
In the first study part, an abdominal phantom was scanned with various extension rings using a CT scanner equipped with a conventional discrete circuit (DC) detector and on the same scanner with an IC detector (120 kVp, 150 effective mAs, and 75 effective mAs). In the second study part, 20 patients were included who underwent abdominal CT both with the IC detector and previously at similar protocol parameters (120 kVp tube current-time product and 150 reference mAs using automated tube current modulation) with the DC detector. Images were reconstructed with filtered back projection.
Results:
Image quality in the phantom was higher for images acquired with the IC compared with the DC detector. There was a gradually increasing noise reduction with increasing phantom sizes, with the highest (37% in the largest phantom) at 75 effective mAs (p < 0.001). In patients, noise was overall significantly (p = 0.025) reduced by 6.4% using the IC detector. Similar to the phantom, there was a gradual increase in noise reduction to 7.9% in patients with a body mass index of 25 kg/m(2) or lower (p = 0.008). Significant correlation was found in patients between noise and abdominal diameter in DC detector images (r = 0.604, p = 0.005), whereas no such correlation was found for the IC detector (r = 0.427, p = 0.060).
Conclusion:
Use of an IC detector in abdominal CT improves image quality and reduces image noise, particularly in overweight and obese patients. This noise reduction has the potential for dose reduction in abdominal CT.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...