Comparative assessment of automated nucleic acid sample extraction equipment for biothreat agents
Warren Vincent Kalina1, Christina Elizabeth Douglas, Susan Rajnik Coyne
1Diagnostic Systems Division, United States Army Medical Research Institute of Infectious Disease, Fort Detrick, Maryland, USA.
Journal of Clinical Microbiology
|January 24, 2014
Abstract:
Magnetic beads offer superior impurity removal and nucleic acid selection over older extraction methods. The performances of nucleic acid extraction of biothreat agents in blood or buffer by easyMAG, MagNA Pure, EZ1 Advanced XL, and Nordiag Arrow were evaluated. All instruments showed excellent performance in blood; however, the easyMAG had the best precision and versatility.