Related Experiment Video
Updated: May 3, 2026

Dynamic Contrast Enhanced Magnetic Resonance Imaging of an Orthotopic Pancreatic Cancer Mouse Model
Published on: April 18, 2015
Modeling and correction of bolus dispersion effects in dynamic susceptibility contrast MRI
Amit Mehndiratta1, Fernando Calamante, Bradley J MacIntosh
1Institute of Biomedical Engineering, University of Oxford, Oxford, United Kingdom.
Purpose:
Bolus dispersion in DSC-MRI can lead to errors in cerebral blood flow (CBF) estimation by up to 70% when using singular value decomposition analysis. However, it might be possible to correct for dispersion using two alternative methods: the vascular model (VM) and control point interpolation (CPI). Additionally, these approaches potentially provide a means to quantify the microvascular residue function.
Methods:
VM and CPI were extended to correct for dispersion by means of a vascular transport function. Simulations were performed at multiple dispersion levels and an in vivo analysis was performed on a healthy subject and two patients with carotid atherosclerotic disease.
Results:
Simulations showed that methods that could not address dispersion tended to underestimate CBF (ratio in CBF estimation, CBFratio = 0.57-0.77) in the presence of dispersion; whereas modified CPI showed the best performance at low-to-medium dispersion; CBFratio = 0.99 and 0.81, respectively. The in vivo data showed trends in CBF estimation and residue function that were consistent with the predictions from simulations.
Conclusion:
In patients with atherosclerotic disease the estimated residue function showed considerable differences in the ipsilateral hemisphere. These differences could partly be attributed to dispersive effects arising from the stenosis when dispersion corrected CPI was used. It is thus beneficial to correct for dispersion in perfusion analysis using this method.
Related Concept Videos
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...
Two-Compartment Open Model: IV Bolus Administration
The disparity between drug input and the sum of drug transfer rates between...
Magnetic Resonance Imaging

