Three-Dimensional Preoperative Virtual Planning in Derotational Proximal Femoral Osteotomy
Published on: February 17, 2023
Accuracy and reliability of linear measurements using 3-dimensional computed tomographic imaging software for Le Fort
Bruno Felipe Gaia1, Lucas Rodrigues Pinheiro1, Otávio Shoite Umetsubo1
1School of Dentistry, University of São Paulo, College of Dentistry, Department of Radiology, Av. Prof. Lineu Prestes, 2227, CEP 05508-900 São Paulo, SP, Brazil.
Abstract:
Our purpose was to compare the accuracy and reliability of linear measurements for Le Fort I osteotomy using volume rendering software. We studied 11 dried skulls and used cone-beam computed tomography (CT) to generate 3-dimensional images. Linear measurements were based on craniometric anatomical landmarks that were predefined as specifically used for Le Fort I osteotomy, and identified twice each by 2 radiologists, independently, using Dolphin imaging version 11.5.04.35. A third examiner then made physical measurements using digital calipers. There was a significant difference between Dolphin imaging and the gold standard, particularly in the pterygoid process. The largest difference was 1.85mm (LLpPtg L). The mean differences between the physical and the 3-dimensional linear measurements ranged from -0.01 to 1.12mm for examiner 1, and 0 to 1.85mm for examiner 2. Interexaminer analysis ranged from 0.51 to 0.93. Intraexaminer correlation coefficients ranged from 0.81 to 0.96 and 0.57 to 0.92, for examiners 1 and 2, respectively. We conclude that the Dolphin imaging should be used sparingly during Le Fort I osteotomy.