Evaluation of tracheal cuff pressure variation in spontaneously breathing patients
Gustavo A Plotnikow1, Nicolas Roux2, Viviana Feld2
1Respiratory Care Services, Clínica Basilea, Argentina ; Intensive Care Unit, Sanatorio Anchorena, Buenos Aires, Argentina.
Background:
Most of the studies referring cuff tubes' issues were conducted on intubated patients. Not much is known about the cuff pressure performance in chronically tracheostomized patients disconnected from mechanical ventilation.
Objective:
To evaluate cuff pressure (CP) variation in tracheostomized, spontaneously breathing patients in a weaning rehabilitation center.
Materials And Methods:
Experimental setup to test instruments in vitro, in which the gauge (TRACOE) performance at different pressure levels was evaluated in six tracheostomy tubes, and a clinical setupin which CP variation over 24 h, every 4 h, and for 6 days was measured in 35 chronically tracheostomized clinically stable, patients who had been disconnected from mechanical ventilation for at least 72 h. The following data were recorded: Tube brand, type, and size; date of the tube placed; the patient's body position; the position of the head; axillary temperature; pulse and respiration rates; blood pressure; and pulse oximetry.
Results:
In vitro difference between the initial pressure (IP) and measured pressure (MP) was statistically significant (P < 0.05). The difference between the IP and MP was significant when selecting for various tube brands (P < 0.05). In the clinical set-up, 207 measurements were performed and the CP was >30 cm H2O in 6.28% of the recordings, 20-30 cm H2O in 42.0% of the recordings, and <20 cm H2O in 51.69% of the recordings.
Conclusion:
The systematic CP measurement in chronically tracheostomized, spontaneously breathing patients showed high variability, which was independent of tube brand, size, type, or time of placement. Consequently, measurements should be made more frequently.
Related Concept Videos
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs...
Physical Assessment of the Respiratory Tract II: Palpation
Thoracic Palpation
Thoracic palpation detects tenderness, masses, lesions, respiratory excursions, and vocal fremitus. The nurse assesses...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like...
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration...