Single sweep three-dimensional carotid ultrasound: reproducibility in plaque and artery volume measurements
Hayrapet Kalashyan1, Ashfaq Shuaib2, Patrick H Gibson3
1Mazankowski Alberta Heart Institute, University of Alberta Hospital, Edmonton, Canada; Stroke Program, Division of Neurology, University of Alberta Hospital, Edmonton, Canada.
Background:
There is a need for non-invasive and accurate techniques for assessment of severity of atherosclerotic disease in the carotid arteries. Recently an automated single sweep three-dimensional ultrasound (3D US) technique became available. The aims of this study were to evaluate the feasibility and reproducibility of the automated single sweep method in a cohort of patients undergoing clinically indicated carotid ultrasound.
Methods:
Consecutive patients with a history of stroke or transient ischemic attack (TIA) and having a plaque in the internal carotid artery (ICA) were recruited for this study. Imaging was performed using a Philips iU 22 ultrasound system equipped with the single sweep volumetric transducer vL 13-5. Analysis was performed offline with software provided by the manufacturer. Two independent observers performed all measurements.
Results:
Of 137 arteries studied (from 79 patients), plaque and artery volumes could be measured in 106 (77%). Reproducibility of plaque volume measurements was assessed in 82 arteries. Bland-Altman analysis demonstrated good inter-observer reproducibility with limits of agreement -0.06 to +0.07 ml. The mean percentage difference between two observers was 5.6% ± 6.0%. Reproducibility of artery volume measurement was assessed in 31 cases. Bland-Altman analysis demonstrated limits of agreement from -0.15 to +0.15 ml. The mean percentage difference was 6.4 ± 5.9%.
Conclusion:
The new automated single sweep 3D ultrasound is feasible in the majority of patients. Good reproducibility in plaque and artery volume measurements makes this technique suitable for serial assessment of carotid plaques.
More Related Videos
09:36A Magnetic Resonance Imaging-based Computational Protocol for Analysis of Plaque Morphology and Hemodynamics in Patients with Carotid Artery Stenosis
Published on: August 12, 2025
10:41Analysis of Cerebral Vasospasm in a Murine Model of Subarachnoid Hemorrhage with High Frequency Transcranial Duplex Ultrasound
Published on: June 3, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System IV: CMRI