A Basic Positron Emission Tomography System Constructed to Locate a Radioactive Source in a Bi-dimensional Space
Published on: February 1, 2016
Influence of positron emitters on standard γ-camera imaging
Lars Jødal1, Pia Afzelius, Svend Borup Jensen
1Department of Nuclear Medicine, Aalborg University Hospital, Aalborg, Denmark; and.
Unlabelled:
Combined PET and SPECT scanning can give supplementary information. However, activity from PET radionuclides can cause background counts and increased dead time in γ camera imaging (SPECT or planar) because the 511-keV photons can penetrate collimators designed for lower energies. This study investigated how to manage this issue, including what levels of PET radionuclides can be tolerated when a γ-camera investigation is performed.
Methods:
Different combinations of (68)Ga (PET radionuclide), (99m)Tc (low-energy radionuclide), and (111)In (medium-energy radionuclide) were scanned by a γ camera. Standard low-, medium-, and high-energy collimators were used with the γ camera. Dead time and counts near and distant from the sources were recorded.
Results:
Down scatter from 511 keV can give rise to a considerable number of counts within the (99m)Tc or (111)In energy windows, especially when the PET source is close to the camera head. Over the full camera head, the PET source can result in more counts per megabecquerel than the SPECT source ((99m)Tc or (111)In). Counts from the PET source were distributed over a large region of the camera head. With medium- and high-energy collimators, the sensitivity to the PET radionuclide was found to be about 10% of the sensitivity to (99m)Tc and about 20% of the sensitivity to (111)In, as measured within a 3-cm-radius region of interest.
Conclusion:
If PET radionuclides of activity 1 MBq or higher are present in the patient at the time of SPECT, a medium-energy collimator should be used. Counts from PET sources will in SPECT usually be seen as a diffuse background rather than as foci. The thick septa of high-energy collimators may result in structure in the image, and a high-energy collimator is recommended only if PET activity is greater than 10 MBq.
More Related Videos
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
X-ray Imaging