Iterative image reconstruction techniques for CT coronary artery calcium quantification: comparison with traditional
Andreas Schindler1, Rozemarijn Vliegenthart, U Joseph Schoepf
1From the Department of Radiology and Radiological Science (A.S., R.V., U.J.S., P.B., U.E., Y.J.C., C.F., P.A.) and Division of Cardiology (U.J.S., U.E.) Medical University of South Carolina, Ashley River Tower, 25 Courtenay Dr, MSC 226, Charleston, SC 29401; Institute of Clinical Radiology, Ludwig-Maximilian-University Hospital, Munich, Germany (A.S., T.S., F.B., K.N.); Center for Medical Imaging-North East Netherlands (R.V.) and Department of Radiology (R.V.), University of Groningen, University Medical Center Groningen, Groningen, the Netherlands; Department of Diagnostic Radiology, University Hospital Freiburg, Freiburg, Germany (P.B.); Department of Cardiology and Intensive Care Medicine, Heart Center Munich-Bogenhausen, Munich, Germany (U.E.); Siemens Healthcare, CT Division, Forchheim, Germany (T.A., S.V., R.R.); and Institute of Clinical Radiology and Nuclear Medicine, University Medical Center Mannheim, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany (C.F., P.A.).
Purpose:
To investigate in vitro and in vivo the use of image-based and raw data-based iterative reconstruction algorithms for quantification of coronary artery calcium by using the Agatston score and subsequent cardiac risk stratification.
Materials And Methods:
In vitro data were obtained by using a moving anthropomorphic cardiac phantom containing calcium inserts of different concentrations and sizes. With institutional review board approval and HIPAA compliance, coronary calcium imaging data of 110 consecutive patients (mean age ± standard deviation, 58.2 years ± 9.8; 48 men) were reconstructed with filtered back projection (FBP), iterative reconstruction in image space (IRIS), and sinogram-affirmed iterative reconstruction (SAFIRE). Image noise was measured and the Agatston score was obtained for all reconstructions. Assignment to Agatston scores and percentile-based cardiac risk categories was compared. Statistical analysis included the Cohen κ coefficient and Friedman and Wilcoxon testing.
Results:
In vitro, mean Agatston scores ± standard deviation for FBP (638.9 ± 9.6), IRIS (622.7 ± 15.2), and SAFIRE (631.4 ± 17.6) were comparable (P = .30). The smallest phantom calcifications were more frequently detected when iterative reconstruction techniques were used. The Agatston scores in the patient cohort were not significantly different among FBP, IRIS, and SAFIRE in paired comparisons (median Agatston score [25th and 75th percentiles]: 76.0 [20.6, 243.9], 76.4 [22, 249.3], and 75.7 [21.5, 49.1], respectively; P = .20 each). Comparison of categorization based on Agatston score percentiles showed excellent agreement for both IRIS and SAFIRE with FBP (κ = 0.975 [0.942-1.00] and κ = 0.963 [0.922-1.00], respectively). The mean effective dose was 1.02 mSv ± 0.51. Mean image noise was significantly (P < .001) higher with FBP than that with iterative reconstructions.
Conclusion:
In comparison with FBP, iterative reconstruction techniques do not have a profound effect on the reproducible quantification of coronary calcium according to Agatston score and subsequent cardiac risk classification, although risk reclassification may occur in a small subset of subjects.
More Related Videos
10:16Retrograde Perfusion and Filling of Mouse Coronary Vasculature as Preparation for Micro Computed Tomography Imaging
Published on: February 10, 2012
12:32Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans
Published on: September 27, 2020
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography