Related Experiment Video
Updated: May 3, 2026

How to Administer Near-Infrared Spectroscopy in Critically ill Neonates, Infants, and Children
Published on: August 19, 2020
Comparison of neonatal skin sensor temperatures with axillary temperature: does skin sensor placement really matter?
Dorothea Schafer1, Sheri Boogaart, Lynette Johnson
1Neonatal Services, Helen DeVos Children's Hospital, Spectrum Health Medical Center, Grand Rapids, Michigan (Mss Schafer, Boogaart, Johnson, Keezel, and Ruperts); and Nursing Practice & Development, Spectrum Health Hospitals, Grand Rapids, Michigan (Dr Vander Laan).
Purpose:
Appropriate thermoregulation affects both morbidity and mortality in the neonatal setting. Nurses rely on information from temperature sensors and radiant warmers or incubators to appropriately maintain a neonate's body temperature. Skin temperature sensors must be repositioned to prevent skin irritation and breakdown. This study addresses whether there is a significant difference between skin sensor temperature readings from 3 locations on the neonate and whether there is a significant difference between skin sensor temperatures compared with digital axillary temperatures.
Subjects:
The study participants included 36 hemodynamically stable neonates, with birth weight of 750 g or more and postnatal age of 15 days or more, in a neonatal intensive care unit. Gestational age ranged from 29.6 to 36.1 weeks at the time of data collection.
Design:
A method-comparison design was used to evaluate the level of agreement between skin sensor temperatures and digital axillary thermometer measurements.
Methods:
When the neonate's skin sensor was scheduled for routine site change, 3 new skin sensors were placed-1 each on the right upper abdomen, left flank, and right axilla. The neonate was placed in a supine position and redressed or rewrapped if previously dressed or wrapped. Subjects served as their own controls, with temperatures measured at all 3 skin sensor sites and followed by a digital thermometer measurement in the left axilla. The order of skin sensor temperature measurements was randomly assigned by a computer-generated number sequence.
Main Outcome Measures:
An analysis of variance for repeated measures was used to test for statistical differences between the skin sensor temperatures. The difference in axillary and skin sensor temperatures was calculated by subtracting the reference standard temperature (digital axillary) from the test temperatures (skin temperatures at 3 different locations), using the Bland-Altman method. The level of significance was set at P < .05.
Principal Results:
No statistically significant differences were found between skin temperature readings obtained from the 3 sites (F2,70 = 2.993, P = .57). Differences between skin temperature readings and digital axillary temperature were also not significant when Bland-Altman graphs were plotted.
Conclusions:
For hemodynamically stable neonates in a supine position, there were no significant differences between skin sensor temperatures on abdomen, flank, or axilla or between skin sensor temperatures and a digital axillary temperature. This may increase nurses' confidence that various sites will produce accurate temperature readings.
More Related Videos
04:34Mouse Body Temperature Measurement Using Infrared Thermometer During Passive Systemic Anaphylaxis and Food Allergy Evaluation
Published on: September 14, 2018
08:22Measuring Cardiac Autonomic Nervous System ANS Activity in Toddlers - Resting and Developmental Challenges
Published on: February 25, 2016
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.