Ultrasound measurement for abdominal aortic aneurysm screening: a direct comparison of the three leading methods
K W H Chiu1, L Ling1, V Tripathi2
1Department of Radiology, Hull and East Yorkshire NHS Trust, Hull, UK.
Objectives:
Ultrasound (US) is non-invasive and cost-effective for screening abdominal aortic aneurysms (AAAs) but there is no universally accepted method to measure the aortic diameter. This study evaluates the accuracy, reproducibility, and repeatability of three methods: inner-to-inner (ITI), leading-to-leading edge (LTL), and outer-to-outer (OTO). The secondary objective of this study was to determine whether aneurysm size or grade of operator had any effect on either intra- or inter-observer variability.
Methods:
Fifty static US images were measured by six assessors (2 vascular radiologists, 2 interventional radiology trainees, and 2 sonographers) on two separate occasions 6 weeks apart. Repeatability and reproducibility were calculated and compared with computed tomography (CT) as the gold standard.
Results:
All three methods have high repeatability and reproducibility when static images are used. The inter-observer reproducibility coefficients between assessors were 0.48 cm, 0.35 cm, and 0.34 cm for ITI, LTL and OTO, respectively. The intra-observer repeatability coefficients between assessors were 0.30 cm, 0.20 cm, and 0.19 cm for ITI, LTL and OTO, respectively. The mean difference between CT and OTO, LTL, and ITI was 1 mm, 3 mm, and 5 mm, respectively (all underestimations) (p < .0001).
Conclusions:
US consistently underestimates aortic size when compared with CT, with ITI demonstrating the greatest underestimation (on average 5 mm). In the UK, this underestimation by the NHS Abdominal Aortic Aneurysm screening programme reduces the sensitivity of the screening test and may impact on the way in which vascular specialists interpret the findings of the screening programme.
More Related Videos
Related Concept Videos
Aneurysm II: Clinical Manifestations and Diagnostic Studies
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...