TEG® and RapidTEG® are unreliable for detecting warfarin-coagulopathy: a prospective cohort study
C Michael Dunham1, Charlene Rabel, Barbara M Hileman
1Trauma/Critical Services, St, Elizabeth Health Center, 1044 Belmont Avenue, Youngstown, OH 44501, USA. dunham.michael@sbcglobal.net.
Background:
Thromboelastography® (TEG) utilizes kaolin, an intrinsic pathway activator, to assess clotting function. Recent published studies suggest that TEG results are commonly normal in patients receiving warfarin, despite an increased International Normalized Ratio (INR). Because RapidTEG™ includes tissue factor, an extrinsic pathway activator, as well as kaolin, we hypothesized that RapidTEG would be more sensitive in detecting a warfarin-effect.
Methods:
Included in this prospective study were 22 consecutive patients undergoing elective cardioversion and receiving warfarin. Prior to cardioversion, blood was collected to assess INR, Prothrombin Time, TEG, and RapidTEG.
Results:
INR Results: 2.8 ± 0.5 (1.6 to 4.2). Prothrombin Time Results: 19.1 ± 2.2 (13.9. to 24.3).TEG Results (Reference Range): R-Time: 8.3 ± 2.7 (2-8); K-Time: 2.1 ± 1.4 (1-3); Angle: 62.5 ± 10.3 (55-78); MA: 63.2 ± 10.3 (51-69); G: 9.4 ± 3.5 (4.6-10.9); R-Time within normal range: 10 (45.5%) with INR 2.9 ± 0.3; Correlation coefficients for INR and each of the 5 TEG variables were insignificant (P > 0.05).RapidTEG Results (Reference Range): ACT: 132 ± 58 (86-118); K-Time: 1.2 ± 0.5 (1-2); Angle: 75.4 ± 5.2 (64-80); MA: 63.4 ± 5.1 (52-71); G: 8.9 ± 2.0 (5.0-11.6); ACT within normal range: 9 (40.9%) with INR 2.7 ± 0.5; Correlation coefficients for INR and each of the 5 RapidTEG variables were insignificant (P > 0.05).
Conclusions:
TEG, using kaolin activation, and RapidTEG, with kaolin and tissue factor activation, were normal in a substantial percent of warfarin patients, despite an increased INR. The false-negative rate for detecting warfarin coagulopathy with either test is unacceptable. The lack of correlation between INR and all TEG and RapidTEG components further indicates that these methodologies are insensitive to warfarin effects. Findings suggest that intrinsic pathway activation may mitigate detection of an extrinsic pathway coagulopathy.
Related Concept Videos
Anticoagulant Drugs: Vitamin K Antagonists and Direct Oral Anticoagulants
Warfarin, a prominent vitamin K antagonist family member, exerts its effect by inhibiting the enzyme VKORC1 (vitamin K epoxide reductase complex 1). By hindering this enzyme, warfarin...
Anticoagulant Drugs: Low-Molecular-Weight Heparins
Therapeutic Drug Monitoring: Drug Analysis Methods
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Therapeutic Drug Monitoring: Affecting Factors
Acute Coronary Syndrome III: Diagnostic Studies