Validation of CT doses of SPECT/CT and PET/CT hybrid devices: lessons learned
Terez Sera1, Tamas Porubszky, Miklos Papos
1aDepartment of Nuclear Medicine, University of Szeged bEuromedic Diagnostics, Szeged c'Frédéric Joliot-Curie' National Research Institute for Radiobiology and Radiohygiene, Budapest, Hungary.
Abstract:
The aim of the study was to check the validity of computed tomographic (CT) doses exhibited by SPECT/CT and PET/CT hybrid devices. Dose measurements were taken from four SPECT/CT and four PET/CT cameras commercially available from different manufacturers. A calibrated ionization chamber was placed in whole-body or head phantoms for the acquisition of CT images with clinically used parameters. Computed tomography dose index (CTDIvol) values were calculated according to the IEC 60601-2-44:1999 formula. The measured CTDIvol doses were compared with those preprogrammed by the manufacturer. In the case of the whole-body phantom, the differences between the measured and displayed values varied between -31 and +24% [European document RP162 (2012) sets up the limit for acceptance criterion as ±20%]. The head phantom data showed either an agreement between -10 and +24%, or an underestimation by two-fold. The latter seemed to be because, while preprogramming the doses, the manufacturer had used the whole-body phantom instead of a proper head phantom. The results of the work demonstrate the need for individual dosimetric calibration of every single X-ray tube. Dosimetric checks should be included in the regular quality control programmes of the SPECT/CT and PET/CT devices. Special attention should be paid to head-and-neck and paediatric protocols, in which the use of a head phantom is recommended for dose calibration.
More Related Videos
09:55Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
10:28Gene Regulation and Targeted Therapy in Gastric Cancer Peritoneal Metastasis: Radiological Findings from Dual Energy CT and PET/CT
Published on: January 22, 2018
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography