Summary
This study shows that 3D fluoroscopic navigation with CT planning is accurate for percutaneous iliosacral screw insertion in pelvic fractures. The system is reliable and clinically feasible for surgeons.
Area of Science:
- Orthopedic surgery
- Medical imaging
- Surgical navigation
Background:
- Percutaneous iliosacral screw insertion demands significant surgical expertise and anatomical knowledge.
- Existing navigation systems still present challenges in achieving precise entry points and trajectories.
- Three-dimensional (3D) fluoroscopic navigation combined with computed tomography (CT)-based planning offers a potential solution.
Purpose of the Study:
- To evaluate the accuracy of 3D fluoroscopic navigation for iliosacral screw placement in displaced cadaveric pelvic fractures.
- To report the technical and clinical outcomes of using CT-3D-fluoroscopy matching navigation for percutaneous iliosacral screw insertion.
Main Methods:
- Simulated posterior pelvic ring disruptions with varying vertical displacements in cadaveric pelvic rings.
- Assessed target registration error using fiducial markers and fluoroscopic imaging.
- Applied CT-3D-fluoroscopy matching navigation for percutaneous iliosacral screw placement in six patients with pelvic ring fractures (AO-OTA types B2 and C1).
Main Results:
- Mean target registration error over the sacrum was 1.2mm, unaffected by fracture type or displacement.
- All 12 inserted iliosacral screws were accurately positioned in the clinical series.
- No postoperative complications, such as nerve palsy, were observed.
Conclusions:
- The CT-3D-fluoroscopy matching navigation system demonstrates accuracy and robustness across various pelvic fracture types and displacements.
- Percutaneous iliosacral screw insertion using this navigation system is clinically feasible and safe.