National comparison of activity measurements of ¹³¹I, ⁶⁰Co, and ¹³³Ba in Indonesia
Related Experiment Video
Updated: May 3, 2026

Enhancing Efficiency and Radiolabeling Yields of Carbon-11 Radioligands for Clinical Research Using the Loop Method
Published on: December 20, 2024
Gatot Wurdiyanto1, Pujadi Marsoem1, Susilo Widodo1
1PTKMR-BATAN, Department of Metrology Radiation, Jalan Lebak Bulus Raya No. 49, PO Box 7043 JKSKL, Jakarta Selatan, DKI Jakarta 12070, Indonesia.
Abstract:
National comparisons of radioactivity measurements of (131)I and (60)Co (in 2010) and of (133)Ba (in 2011) were carried out within the framework of the National Radiation Metrology Laboratory Program of Indonesia, coordinated by PTKMR-BATAN. Eleven laboratories took part in the comparison, and all measurements were made using gamma spectrometry, on point sources with an activity between 2000Bq and 6000Bq. Several laboratories reported values which were more than 10% different from the reference value for (131)I and (133)Ba. Possible reasons for these differences are discussed. This program will be continued with other radionuclides to maintain and control quality assurance for the local laboratories.