Comparison of the Complior Analyse device with Sphygmocor and Complior SP for pulse wave velocity and central pressure assessment
Related Experiment Video
Updated: May 3, 2026

Measuring the Carotid to Femoral Pulse Wave Velocity Cf-PWV to Evaluate Arterial Stiffness
Published on: May 3, 2018
Francesco Stea1, Erwan Bozec, Sandrine Millasseau
1aInternal Medicine Department, Pisa University bIstituto di Fisiologia Clinica Consiglio Nazionale delle Ricerche, Pisa, Italy cInserm U970, Université Paris Descartes, Assistance Publique hôpitaux de Paris, Paris dPulse Wave Consulting, Saint Leu la Foret, France.
Background:
The Complior device (Alam Medical, France) was used in epidemiological studies which established pulse wave velocity (PWV) as a cardiovascular risk marker. Central pressure is related, but complementary to PWV and also associated to cardiovascular outcomes. The new Complior Analyse measures both PWV and central blood pressure during the same acquisition. The aim of this study was to compare PWV values from Complior Analyse with the previous Complior SP (PWVcs) and with Sphygmocor (PWVscr; AtCor, Australia), and to compare central systolic pressure from Complior Analyse and Sphygmocor.
Method:
Peripheral and central pressures and PWV were measured with the three devices in 112 patients. PWV measurements from Complior Analyse were analysed using two foot-detection algorithms (PWVca_it and PWVca_cs). Both radial (ao-SBPscr) and carotid (car-SBPscr) approaches from Sphygmocor were compared to carotid Complior Analyse measurements (car-SBPca). The same distance and same calibrating pressures were used for all devices.
Results:
PWVca_it was strongly correlated to PWVscr (R(2) = 0.93, P < 0.001) with a difference of 0.0 ± 0.7 m/s. PWVca_cs was also correlated to PWVcs (R(2) = 0.90, P < 0.001) with a difference of 0.1 ± 0.7 m/s. Central systolic pressures were strongly correlated. The difference between car-SBPca and ao-SBPscr was 3.1 ± 4.2 mmHg (P < 0.001), statistically equivalent to the difference between car-SBPscr and ao-SBPscr (3.9 ± 5.8 mmHg, P < 0.001), whilst the difference between car-SBPca and car-SBPscr was negligible (-0.7 ± 5.6 mmHg, P = NS).
Conclusion:
The new Complior Analyse device provides equivalent results for PWV and central pressure values to the Sphygmocor and Complior SP. It reaches Association for the Advancement of Medical Instrumentation standard for central blood pressure and grades as excellent for PWV on the Artery Society criteria. It can be interchanged with existing devices.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Measurement of Blood Pressure