Related Experiment Video
Updated: May 3, 2026

A Two-interval Forced-choice Task for Multisensory Comparisons
Published on: November 9, 2018
A user-operated audiometry method based on the maximum likelihood principle and the two-alternative forced-choice
Jesper Hvass Schmidt1, Christian Brandt, Ellen Raben Pedersen
1* Department of Audiology, Odense University Hospital, University of Southern Denmark , Odense , Denmark.
A new user-operated pure-tone audiometry system using the two-alternative forced-choice (2AFC) method demonstrates high test-retest reliability. This system offers a dependable alternative to traditional audiometry for research, particularly for musicians.
Area of Science:
- Audiology
- Psychoacoustics
- Biomedical Engineering
Background:
- Traditional audiometry requires skilled operators and can be influenced by subjective response criteria.
- Musicians, often involved in research, require reliable hearing assessments.
Purpose of the Study:
- To develop a user-operated pure-tone audiometry method using maximum likelihood (MML) and two-alternative forced-choice (2AFC).
- To achieve high test-retest reliability without external operators and minimize response criterion variability.
- To provide a user-friendly audiometry tool for research, especially for musicians.
Main Methods:
- The study evaluated the test-retest reliability of the developed user-operated audiometry system.
- A comparison was made between the user-operated system and traditional audiometry.
- 38 naive listeners participated in the reliability testing, and 41 subjects were involved in the comparison study.
Main Results:
- The user-operated 2AFC audiometry showed repeatability comparable to traditional methods, with standard deviations of differences ranging from 3.9 dB to 5.2 dB across 250-8000 Hz.
- Hearing thresholds obtained via user-operated 2AFC audiometry were slightly lower (1-2 dB) at most frequencies compared to traditional audiometry.
Conclusions:
- User-operated 2AFC audiometry requires no specialized skills and offers acceptable, reliable repeatability similar to traditional audiometry.
- This user-operated system is a viable and reliable alternative to conventional audiometry for research applications.
More Related Videos
14:05Behavioral Assessment of Hearing in 2 to 4 Year-old Children: A Two-interval, Observer-based Procedure Using Conditioned Play-based Responses
Published on: January 23, 2017
07:28Psychophysically-anchored, Robust Thresholding in Studying Pain-related Lateralization of Oscillatory Prestimulus Activity
Published on: January 21, 2017