[Noninvasive blood pressure measurement using high definition oscillometry in horses with heart diseases]
1Prof. Dr. Heidrun Gehlen, Dipl. ECEIM, Klinik für Pferde, Allgemeine Chirurgie und Radiologie, Fachbereich Veterinärmedizin, Freie Universität Berlin, Oertzenweg 19b, 14163 Berlin,
Objective:
This study evaluated a device for blood pressure measurement using high definition oscillometry (HDO) for its acceptability and practicability in the horse. Furthermore, we evaluated whether this device can determine differences in blood pressure between healthy horses and horses with heart diseases and between different measurement times.
Material And Methods:
The study population consisted of 60 horses (aged 4-31 years). All horses underwent clinical examination, conventional echocardiography (including cardiac output measurement at the pulmonary artery), repeated noninvasive blood pressure measurement at the coccygeal artery (measurement of systolic [SAP], diastolic [DAP] and mean [MAP] arterial blood pressure) and Doppler sonographic measurement of arterial blood flow at the lateral digital artery. Thirty horses were considered to be cardiologically healthy based upon previous examination. In 30 horses, various heart diseases (mitral valve insufficiency, aortic valve insufficiency and atrial fibrillation) were diagnosed. The diurnal variations in blood pressure in 18 healthy horses and 18 horses with heart disease were determined in the morning and evening. In three horses suffering from atrial fibrillation, the blood pressure was determined before and after successful cardioversion.
Results:
In healthy horses the following mean blood pressures were found: SAP 118 ± 21 mmHg, DAP 70 ± 19 mmHg and MAP 87 ± 19 mmHg. No significant difference was found between healthy horses and horses with heart diseases. An insufficiency of the aortic valve tended to result in a higher SAP (p = 0.05), whereas atrial fibrillation lowered blood pressure compared to healthy horses. With respect to the diurnal rhythm of the blood pressure, healthy horses displayed a slight increase in the systolic arterial pressure in the evening.
Conclusion:
The HDO-sphygmomanometer facilitates a straightforward, indirect way to determine blood pressure that delivers meaningful values. However, the method is unsuited to monitor the development or therapy of cardiac diseases.
More Related Videos
09:17High-Resolution Endocardial and Epicardial Optical Mapping in a Sheep Model of Stretch-Induced Atrial Fibrillation
Published on: July 29, 2011
11:50High-frequency High-resolution Echocardiography: First Evidence on Non-invasive Repeated Measure of Myocardial Strain, Contractility, and Mitral Regurgitation in the Ischemia-reperfused Murine Heart
Published on: July 9, 2010
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)