Investigation of potential interferences on the measurement of dissolved reactive phosphate using zirconium
Qin Sun1, Yifei Chen2, Di Xu3
1Key Laboratory of Integrated Regulation and Resource Development on Shallow Lakes, Ministry of Education, College of Environment, Hohai University, Nanjing 210098, China. sunqinnj@hhu.edu.cn
Abstract:
A diffusive gradients in thin films (DGT) technique based on hydrous zirconium oxide (Zr-oxide) has been recently developed for the measurement of dissolved reactive phosphate (DRP). In this study, the detailed performance of the DGT technique is reported. Spiking experiments revealed that several orthophosphate monoester compounds contributed to the Zr-oxide DGT measurements of DRP. However, such a phenomenon is unlikely to occur during field conditions due to the low concentration of organic P in typical natural waters. The presence of Cl- (up to 106 g/L), SO4(2-) (up to 16 g/L), HCO3- (up to 817 g/L), and AsO2- and AsO4(3-) (both up to 1 mg As/L) in solutions had negligible effects on the measurement of DRP. The threshold concentrations of Cl-, SO4(2-) and HCO3- have been increased from previous reports for the measurements of DRP using other adsorbent-based DGT techniques. The capacity for DGT measurements of DRP decreased with increasing solution pH (4.2-9.2). The lowest capacity (95 microg P/cm2 at pH 9.2) was still greater than that of other DGT techniques that are usually used for the measurement of DRP (2-12 microg P/cm2). The Zr-oxide binding gel could be stored for up to 2 years without any aging effect. This period of validity was considerably longer than the ferrihydrite binding gel that is commonly used in present DGT devices (6 months). The field application revealed that the concentrations of DRP measured in three fresh water samples using the Zr-oxide DGT technique were in agreement with those of the traditional colorimetric method.
More Related Videos
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...