Summary
Transcatheter aortic valve implantation (TAVI) exposes patients to higher radiation doses than coronary angiography. Heart team members receive significant radiation, necessitating enhanced safety protocols for TAVI procedures.
Area of Science:
- Cardiovascular Interventions
- Medical Radiation Physics
- Patient Safety
Background:
- Transcatheter aortic valve implantation (TAVI) is a crucial treatment for high-risk aortic stenosis patients.
- Radiation exposure during TAVI procedures for patients and staff is not well-documented.
Purpose of the Study:
- To quantify patient and heart team radiation doses during transapical TAVI (TA-TAVI).
- To compare these doses with established radiation safety limits.
Main Methods:
- Dose measurements using thermoluminescence and film dosimeters on 22 TA-TAVI patients.
- Dosimeters placed on seven body parts for surgeons, cardiologists, anesthesiologists, assistants, and nurses.
- Comparison with international recommended radiation dose limits.
Main Results:
- Mean radiation time was 6.1 minutes; mean patient dose-area product was 8.661 µGy·m².
- Effective doses per intervention ranged from 0.001 mSv (nurse) to 0.05 mSv (assistant).
- Maximum hand doses reached 0.7 mSv; lower leg and genital doses were also recorded.
Conclusions:
- TA-TAVI patients receive higher X-ray doses than during coronary interventions.
- Radiation exposure for heart team members after 100 procedures approximates an interventional cardiologist's annual dose.
- Optimal X-ray protection and safety measures for patients and staff are critical and require re-evaluation.
Keywords:
Aortic valveRadiationSafetyTransapical aortic valve implantationTranscatheter aortic valve implantationValvesX-ray