Occurrence of longitudinal stent compression before stent deployment: Two case studies.
Adel Aminian1, Jacques Lalmand1, Dariouch Dolatabadi1
1Adel Aminian, Jacques Lalmand, Dariouch Dolatabadi, Division of Cardiology, Centre Hospitalier Universitaire de Charleroi, 6000 Charleroi, Belgium.
World Journal of Cardiology
|February 15, 2014
Summary
New stent designs can cause longitudinal stent deformation (LSD) before deployment. This occurs when the guide catheter crushes the stent edge, a complication requiring specific management strategies.
Area of Science:
- Cardiovascular Interventions
- Medical Device Engineering
Background:
- Longitudinal stent deformation (LSD) is increasingly reported post-deployment.
- Few studies have documented LSD occurring prior to stent deployment.
- Newer stent designs feature fewer connectors, potentially reducing longitudinal strength.
Purpose of the Study:
- To describe cases of pre-deployment LSD.
- To identify risk factors and management strategies for this complication.
Main Methods:
- Case report analysis of two instances of pre-deployment LSD.
- Review of stent design modifications and their impact on mechanical properties.
- Discussion of risk factors and procedural management.
Main Results:
- Two cases of LSD occurred prior to deployment due to guide catheter interaction with the crimped stent.
- Stent length and design modifications are identified as contributing factors.
- Effective management strategies involve technical guidelines and fluoroscopic guidance.
Conclusions:
- Pre-deployment LSD is a previously unrecognized complication linked to modern stent designs.
- Awareness of risk factors and adherence to specific procedural techniques are crucial for prevention and management.
Keywords:
ComplicationLongitudinal stent compressionLongitudinal stent deformationPercutaneous coronary interventionStent structure