Unique hard scleral lens post-LASIK ectasia fitting.
Rajeswari Mahadevan1, Divya Jagadeesh, Rajni Rajan
1*PhD †MPhil(Optom) Contact Lens Department, Medical Research Foundation, Sankara Nethralaya, Chennai, India (all authors).
Summary
Prosthetic Replacement of Ocular Surface Ecosystem (PROSE) devices can help patients with post-LASIK corneal ectasia. Scleral toricity impacts PROSE device fit and function, requiring careful consideration for optimal vision correction.
Area of Science:
- Ophthalmology
- Biomedical Engineering
Background:
- Corneal ectasia, including post-LASIK, involves progressive corneal thinning and steepening, leading to vision impairment.
- Prosthetic Replacement of Ocular Surface Ecosystem (PROSE) devices, a type of scleral lens, are vital for vision salvage when other treatments fail.
- Patient symptoms include glare, ghosting, and reduced visual acuity.
Observation:
- A case report details the use of a PROSE device in an Indian patient with post-LASIK corneal ectasia.
- The patient presented with a larger palpebral aperture, steeper cornea, and significant scleral toricity.
- A smaller-diameter PROSE device with a flatter haptic was chosen due to scleral toricity, deviating from the typical recommendation for steeper corneas.
Findings:
- Scleral toricity can influence PROSE device vault and haptic fit, potentially altering the corneal-contact lens relationship.
- Even with appropriate device selection, scleral toricity may contribute to persistent symptoms after extended wear.
- Adaptation to the PROSE device can lead to changes in vault due to haptic settling on the sclera.
Implications:
- Understanding the scleral profile is crucial for optimizing PROSE device fitting and patient outcomes in complex cases.
- This case highlights the need for individualized PROSE device selection considering not only corneal but also scleral topography.
- Further research into scleral topography's impact on scleral lens wear is warranted to improve management of corneal ectasia.