Effect of reference point on visual evoked potentials: clinical relevance.
P Y Shih1, M J Aminoff, D S Goodin
1Department of Neurology, University of California, San Francisco 94143.
Electroencephalography and Clinical Neurophysiology
|July 1, 1988
Summary
Using linked mastoids as a reference for visual evoked potential (VEP) recordings can reduce interpretive errors compared to vertex or mid-frontal references. However, this method may yield more technically inadequate recordings due to smaller P100 amplitudes and muscle artifact.
Area of Science:
- Neuroscience
- Ophthalmology
- Clinical Electrophysiology
Background:
- Visual evoked potential (VEP) recordings are crucial for assessing the visual pathway.
- Current clinical practice often uses mid-occipital recording referenced to the vertex or mid-frontal region.
- Potential interpretive errors can arise from non-ideal reference electrode placement.
Purpose of the Study:
- To evaluate the utility of using linked mastoids as a reference electrode in VEP recordings.
- To compare the diagnostic yield and technical quality of VEPs recorded with linked mastoid versus conventional references.
- To determine if linked mastoid referencing can reduce interpretive errors in VEP analysis.
Main Methods:
- Simultaneous VEP recordings were performed using both conventional (mid-occipital to vertex/mid-frontal) and linked mastoid reference montages.
- VEPs were analyzed for diagnostic yield, waveform clarity, and technical adequacy.
- Factors contributing to technically inadequate recordings, such as P100 amplitude and muscle artifact, were assessed.
Main Results:
- Both recording derivations demonstrated similar diagnostic yield.
- The linked mastoid reference provided a greater number of technically inadequate recordings.
- Smaller P100 component size and increased muscle artifact contamination were observed with the linked mastoid reference.
Conclusions:
- Linked mastoid referencing can help avoid interpretive errors by using a more inactive reference point.
- Despite similar diagnostic yield, the linked mastoid method may require careful technical execution to mitigate artifact.
- Simultaneous recording with multiple derivations is recommended for clarifying ambiguous VEP waveforms.