Related Experiment Video
Updated: May 2, 2026

Protocols for Testing the Toxicity of Novel Insecticidal Chemistries to Mosquitoes
Published on: February 13, 2019
Interlaboratory testing of Insent e-tongues
Miriam Pein1, Xolani Dereck Gondongwe2, Masaaki Habara3
1Institute of Pharmaceutics and Biopharmaceutics, Heinrich-Heine University Duesseldorf, Universitaetsstr. 1, 40225 Duesseldorf, Germany.
Abstract:
The first interlaboratory testing of electronic taste sensing systems was performed within five participating centers, each working with the Insent (Insent Inc., Atsugi-Shi, Japan) e-tongue. Preparation of the samples for the comprised four experiments, shipping of the samples and evaluation of the results was performed at the University of Duesseldorf. The sensitivity (in this case the difference between lowest and highest sensor response) and slope of the regression line values, obtained within Experiment 1 and 2, have been found to serve as applicable evaluation criterions for interlaboratory comparability. Modified sensor responses could be attributed to aged sensors, but did not influence the results of either Experiment 3, dealing with the evaluation of film formulations, or Experiment 4, dealing with the evaluation of minitablet formulations, in a great amount. Presented PCA Score and Loading Scatter Plots as well as Euclidean distance patterns based on the raw sensor responses confirmed the comparable performance of Insent e-tongues of the participating centers.