Related Experiment Video
Updated: May 2, 2026

Measurement of Tissue Oxygenation Using Near-Infrared Spectroscopy in Patients Undergoing Hemodialysis
Published on: October 2, 2020
Online Hemoglobin and Oxygen Saturation Sensing During Continuous Renal Replacement Therapy with Regional Citrate
Lenar T Yessayan1, Jerry Yee, Stan Frinak
1From the *Division of Nephrology and Hypertension, Henry Ford Hospital, Detroit, Michigan and †Division of Pulmonary and Critical Care Medicine, Henry Ford Hospital, Detroit, Michigan.
Abstract:
Optical hemoglobin and oxygen saturation sensor (OHOS) monitor when used in combination with other hemodynamic tools may be useful for continuous hemodynamic monitoring during ultrafiltration. The stand-alone OHOS monitor can easily be deployed predialyzer into the extracorporeal circuit of continuous renal replacement therapy (CRRT) systems. To maximize the accuracy of the OHOS in 24 hr CRRT systems, clotting in the optical blood chamber and the presensor dilution incurred by replacement fluid should be minimized. Sustained low-efficiency dialysis (SLED) with regional citrate anticoagulation is a therapy that incorporates an OHOS and maintains the overall reliability of hemoglobin (Hb) and saturation sensing. The system operates at a blood flow rate of 60 ml/min and a fixed acid citrate infusion rate of 150 ml/hr. The presensor dilution incurred by concentrated citrate infusion would result in a minimal Hb dilution (<0.7 g/dl) while minimizing optical blood chamber clotting during 24 hr SLED.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
04:44Author Spotlight: Enhancing Vascular Function and Physical Capacity in Cardiovascular Disease Through Novel Interventions and NIRS Technology
Published on: March 22, 2024
Related Concept Videos
Continuous Renal Replacement Therapy
Extracorporeal Removal of Drugs: Continuous Renal Replacement Therapy
Hemodialysis II: Procedure and Complications
Hemodialysis III: Nursing Management
Hemodialysis I: Introduction
Renal Regulation of Acid-Base Balance
In the kidneys, cells within the proximal convoluted tubules (PCT) and the collecting ducts secrete hydrogen ions (H+) into the tubular fluid. Specifically, in the PCT, Na+/H+ antiporters secrete H+ while reabsorbing Na+.