A widely applicable plasma renin activity assay by LC-MS/MS with offline solid phase extraction
Laura J Owen1, Joanne Adaway, Karen Morris
1Biochemistry Department, University Hospital of South Manchester, Manchester, UK.
Annals of Clinical Biochemistry
|February 27, 2014
Summary
This study presents an offline modification for measuring plasma renin activity using LC-MS/MS, offering a practical alternative to specialized equipment. The new method provides comparable results to existing assays, making it accessible to more laboratories.
Area of Science:
- Clinical Chemistry
- Analytical Chemistry
- Mass Spectrometry
Background:
- Plasma renin activity (PRA) measurement is crucial for diagnosing and managing various conditions.
- Radioimmunoassay (RIA) is a traditional method for PRA, but LC-MS/MS offers an attractive alternative.
- Existing LC-MS/MS methods for PRA often require specialized equipment, limiting accessibility.
Purpose of the Study:
- To develop and validate an offline modification for PRA measurement using LC-MS/MS.
- To provide a method that does not require specialized equipment, enhancing accessibility for LC-MS/MS users.
- To compare the performance of the offline method with a previously published online method.
Main Methods:
- Samples were prepared and incubated for 6.5 h and 24 h.
- Offline solid-phase extraction was performed using a commercially available 96-well plate.
- Method comparison was conducted against a previously established online solid-phase extraction assay.
Main Results:
- The offline method yielded results comparable to the online method across the measured concentration range.
- Analytical performance of the offline method was found to be similar to the online assay.
- A 24-hour incubation period was determined to be unnecessary for low-concentration samples.
Conclusions:
- The described offline method is a viable alternative to existing PRA assays.
- This method does not necessitate highly specialized equipment, broadening its applicability.
- The approach is suitable for the majority of laboratories equipped with LC-MS/MS systems.