Atraumaticity study of 2 cochlear implant electrode arrays
Manuel Manrique1, Sebastián Picciafuoco, Raquel Manrique
1*Otorhinolaryngology Department, University Clinic of Navarra; and †Radiology Department, University Clinic of Navarra, Pamplona, Spain.
Summary
This study found that neither the HiFocus 1J nor HiFocus Helix electrode arrays caused significant insertion trauma in human temporal bones. Both electrode arrays demonstrated atraumatic insertion, preserving cochlear anatomy during surgery.
Area of Science:
- Otolaryngology
- Neurosurgery
- Biomedical Engineering
Background:
- Atraumatic surgery and advanced electrode arrays are critical in modern otologic procedures.
- Preserving cochlear anatomy and function is paramount for successful outcomes.
- Morphological evaluation of surgical trauma is essential for refining techniques.
Purpose of the Study:
- To evaluate the atraumaticity of two electrode arrays (HiFocus 1J and HiFocus Helix) in human temporal bones.
- To compare surgical trauma based on morphologic and histologic parameters.
- To assess the safety of electrode array insertion in the cochlea.
Main Methods:
- Ten human temporal bones were divided into two groups for atraumatic insertion of HiFocus 1J and HiFocus Helix electrode arrays.
- Computed tomography and histologic analysis were used for morphologic and histologic evaluation.
- Insertion depth, angle, and location within the cochlea were meticulously analyzed.
Main Results:
- HiFocus 1J: Successful scala tympani insertion in 4/5 specimens with lateral array location.
- HiFocus Helix: Successful scala tympani insertion in 4/5 specimens with modiolar array location.
- Neither array demonstrated significant correlation between dimensions and insertion depth, indicating consistent placement.
Conclusions:
- Both HiFocus 1J and HiFocus Helix electrode arrays are safe for insertion.
- The study concludes that neither electrode array causes significant insertion trauma.
- These findings support the use of these electrode arrays in cochlear implantation surgeries.