The effect of 50 Hz external electrical interference on implanted cardiac pacemakers
G C Kaye1, G S Butrous, A Allen
1Department of Cardiology, St. Bartholomew's Hospital, London, England.
Pacing and Clinical Electrophysiology : PACE
|July 1, 1988
Summary
Injecting 50 Hz alternating current into patients with VVI pacemakers caused interference. Most pacemakers reverted to interference mode at currents between 29-250 microA, affecting pacing function.
Area of Science:
- Biomedical Engineering
- Cardiology
- Electrophysiology
Background:
- Implanted cardiac pacemakers are crucial for managing bradyarrhythmias.
- Understanding pacemaker susceptibility to external electromagnetic interference is vital for patient safety.
- Previous studies have explored various interference sources, but controlled testing methods for 50 Hz AC interference require further investigation.
Purpose of the Study:
- To evaluate the effects of injected 50 Hz alternating current (AC) on the function of VVI pacemakers.
- To determine the current thresholds at which pacemakers exhibit interference or reversion modes.
- To assess the safety and reproducibility of using current injection as a method for testing pacemaker sensitivity.
Main Methods:
- 18 patients with implanted unipolar VVI pacemakers were subjected to injected 50 Hz AC (0-600 microA) via surface electrodes.
- Pacemaker response, including reversion to interference mode and pacing irregularities, was monitored.
- Interference voltage was measured in 10 additional patients using temporary pacing catheters.
- Factors influencing sensitivity, such as electrode placement, patient posture, and respiration, were analyzed.
Main Results:
- Most VVI pacemakers reverted to interference mode within a current range of 29-250 microA.
- Irregular and inappropriate pacing occurred just below the reversion current threshold, varying from 27-246 microA.
- Pacemaker sensitivity was influenced by electrode location, patient posture, and deep inspiration.
- Measured interference voltage also varied with posture and respiration.
Conclusions:
- Injected 50 Hz AC can safely and reproducibly test VVI pacemaker sensitivity to external interference.
- Pacemaker reversion and irregular pacing occur at specific, patient-dependent current levels.
- Patient factors like posture and respiration significantly impact pacemaker response to interference.