Multi-timescale Microscopy Methods for the Characterization of Fluorescently-labeled Microbubbles for Ultrasound-Triggered Drug Release
Published on: June 12, 2021
The delayed onset of subharmonic and ultraharmonic emissions from a phospholipid-shelled microbubble contrast agent
Himanshu Shekhar1, Ivy Awuor1, Keri Thomas2
1Department of Electrical and Computer Engineering, University of Rochester, Rochester, NY, USA.
Abstract:
Characterizing the non-linear response of microbubble contrast agents is important for their efficacious use in imaging and therapy. In this article, we report that the subharmonic and ultraharmonic response of lipid-shelled microbubble contrast agents exhibits a strong temporal dependence. We characterized non-linear emissions from Targestar-p microbubbles (Targeson Inc., San Diego, CA, USA) periodically for 60 min, at 10 MHz excitation frequency. The results revealed a considerable increase in the subharmonic and ultraharmonic response (nearly 12-15 and 5-8 dB) after 5-10 min of agent preparation. However, the fundamental and the harmonic response remained almost unchanged in this period. During the next 50 min, the subharmonic, fundamental, ultraharmonic, and harmonic responses decreased steadily by 2-5 dB. The temporal changes in the non-linear behavior of the agent appeared to be primarily mediated by gas-exchange through the microbubble shell; temperature and prior acoustic excitation based mechanisms were ruled out. Further, there was no measurable change in the agent size distribution by static diffusion. We envisage that these findings will help obtain reproducible measurements from agent characterization, non-linear imaging, and fluid-pressure sensing. These findings also suggest the possibility for improving non-linear imaging by careful design of ultrasound contrast agents.
Related Concept Videos
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...