Dual-source parallel radiofrequency transmission for magnetic resonance breast imaging at 3T: any added clinical
Lei Jiang1, Yiming Zhou2, Cheng Zhou1
1Radiology Department, Beijing Hospital, the Fifth Affiliated College of Peking University, Beijing, China.
Purpose:
To investigate the influence of dual-source parallel radiofrequency (RF) excitation on clinical breast MR images.
Methods:
A 3T MR system with both dual-source and conventional single-source RF excitations was used to examine 22 patients. Axial TSE-T2WI with fat suppression, TSE-T1WI without fat suppression, THRIVE (3D field echo) and DWI (SE-EPI) were obtained by using both excitation techniques. Image homogeneity, image contrast and lesion conspicuity were measured or independently scored by two radiologists and were compared by paired-sample t test or Wilcoxon test.
Results:
Both excitations revealed 24 lesions. For SE sequences using dual-source mode, image homogeneity was improved (P=0.00), scan time was reduced, and ghost artifacts on DWI were significantly reduced (P=0.00). However, image contrast was not increased and lesion conspicuity had no significant difference between two modes, except DWI on which lesion conspicuity was significantly improved (P=0.00), due to less ghost artifacts. For field-echo sequence, image homogeneity, acquisition time, image contrast and lesion conspicuity had no significant difference between the two modes.
Conclusions:
Dual-source parallel RF transmission has some added value for improving breast image quality. However, its value is limited in terms of improving lesion detection and characterization.