Purpose:
To achieve detection of glutamate (Glu), glutamine (Gln), and glutathione (GSH) by minimizing the N-acetyl-aspartate (NAA) multiplet signals at 2.49 ppm using a echo time (TE) -optimized PRESS pulse sequence and a novel J-suppression radiofrequency pulse.
Methods:
Using density matrix simulations, a PRESS sequence with (TE1 , TE2 ) = (69, 37) ms and an inserted 90° J-suppression pulse were found to minimize the NAA multiplet at 2.49 ppm.
Results:
NAA phantom experiments confirmed the successful suppression of the NAA multiplet at 2.49 ppm. A study of eight healthy volunteers found both Glu and Gln to be significantly higher in gray matter (GM) dominant medial prefrontal cortex voxels than in white matter (WM) dominant right frontal cortex voxels. Time-course (1) H spectra acquired during intravenous [U-(13) C6 ]glucose infusion showed gradually changing Glu C4 and Gln C4 proton resonance signals in a spectral pattern predicted by numerical simulations.
Conclusion:
Reliable detection of Glu, Gln, and GSH was achieved. Glu and Gln levels were significantly higher in frontal lobe GM than in frontal lobe WM. It is feasible to use the proposed proton MR spectroscopy method to measure the kinetics of (13) C incorporation into Glu and Gln during infusion of (13) C labeled glucose.
More Related Videos
13:56The Use of Magnetic Resonance Spectroscopy as a Tool for the Measurement of Bi-hemispheric Transcranial Electric Stimulation Effects on Primary Motor Cortex Metabolism
Published on: November 19, 2014
09:30Quantitative Magnetic Resonance Imaging of Skeletal Muscle Disease
Published on: December 18, 2016
Related Concept Videos
Magnetic Resonance Imaging
Insensitive Nuclei Enhanced by Polarization Transfer (INEPT)
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...