Invasive blood pressure recording comparing nursing charts with an electronic monitor: a technical report
Benjamin T Wong1, Neil J Glassford2, Victoria Bion2
1Department of Intensive Care, Austin Hospital, Melbourne, VIC, Australia. rinaldo.bellomo@austin.org.au.
Background:
Blood pressure management (assessed using nursing charts) in the early phase of septic shock may have an effect on renal outcomes. Assessment of mean arterial pressure (MAP) values as recorded on nursing charts may be inaccurate.
Aim:
To determine the difference between hourly blood pressure values as recorded on the nursing charts and hourly average blood pressure values over the corresponding period obtained electronically from the bedside monitor.
Methods:
We studied 20 patients with shock requiring vasopressor support and invasive blood pressure monitoring. Hourly blood pressure measurements were recorded on the nursing charts over a 12-hour period. Blood pressure values recorded every 10 minutes were downloaded from electronic patient monitors over the corresponding period. The hourly average of the 10-minute blood pressure values was compared with the measurements recorded on the nursing charts.
Results:
We assessed 240 chart readings and 1440 electronic recordings. Average chart MAP was 72.54 mmHg and average electronic monitor MAP was 71.54 mmHg. MAP data from the two sources showed a strong correlation (ρ0.71, P < 0.005). Bland-Altman assessment revealed acceptable agreement, with a mean bias of 1mmHg and 95% limits of agreement of -11.76 mmHg and 13.76 mmHg. Using average data over 6 hours, 95% limits of agreement narrowed to -6.79mmHg and 8.79mmHg.
Conclusion:
With multiple measurements over time, mean blood pressure as recorded on nursing charts reasonably approximates mean blood pressure recorded on the monitor.
More Related Videos
05:01A Detailed Protocol for Physiological Parameters Acquisition and Analysis in Neurosurgical Critical Patients
Published on: October 17, 2017
06:51Development of an Algorithm to Perform a Comprehensive Study of Autonomic Dysreflexia in Animals with High Spinal Cord Injury Using a Telemetry Device
Published on: July 29, 2016
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Errors occurring during blood pressure monitoring
Several factors...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pre-Procedural Guidelines for Assessing Blood Pressure