Related Experiment Video
Updated: May 2, 2026

Measuring Carbon-based Contaminant Mineralization Using Combined CO2 Flux and Radiocarbon Analyses
Published on: October 21, 2016
Complementary system for long term measurements of radon exhalation rate from soil
Abstract:
A special set-up for continuous measurements of radon exhalation rate from soil is presented. It was constructed at Laboratory of Radiometric Expertise, Institute of Nuclear Physics Polish Academy of Sciences (IFJ PAN), Krakow, Poland. Radon exhalation rate was determined using the AlphaGUARD PQ2000 PRO (Genitron) radon monitor together with a special accumulation container which was put on the soil surface during the measurement. A special automatic device was built and used to raise and lower back onto the ground the accumulation container. The time of raising and putting down the container was controlled by an electronic timer. This set-up made it possible to perform 4-6 automatic measurements a day. Besides, some additional soil and meteorological parameters were continuously monitored. In this way, the diurnal and seasonal variability of radon exhalation rate from soil can be studied as well as its dependence on soil properties and meteorological conditions.