Fluctuation of an ion beam extracted from an AC filament driven Bernas-type ion source
N Miyamoto1, Y Okajima1, M Wada1
1Graduate School of Science and Engineering, Doshisha University, Kyotanabe, Kyoto 610-0321, Japan.
Abstract:
Argon ion beam fluctuation from an AC filament driven Bernas-type ion source is observed. The ion beam was measured by an 8 measurement elements beam profile monitor. The amplitude of the beam current fluctuation stayed in the same level from 100 Hz to 1 kHz of the filament heating frequency. The beam current fluctuation frequency measured by the beam profile monitor was equal to the frequency of the AC filament operation. The fluctuation amplitudes of the beam current by AC operation were less than 7% and were in the same level of the DC operation.