Related Experiment Video
Updated: May 2, 2026

Cryogenic Liquid Jets for High Repetition Rate Discovery Science
Published on: May 9, 2020
Recondensation performance of liquid helium cryostat for a 28 GHz electron cyclotron resonance ion source
Seyong Choi1, Byoung-Seob Lee1, Jin Yong Park1
1Busan Center, Korea Basic Science Institute, Busan 609-735, South Korea.
Abstract:
Cryostat performance is essential for the stable operation of a superconducting magnet. A closed-cycle liquid helium cryostat was adopted for use for a superconducting electron cyclotron resonance (ECR) ion source by recondensing liquid helium vapor. The goal was to maintain the liquid helium filled reservoir at a constant level without transferring any liquid helium during the normal operation of the ECR ion source. To accomplish this, Gifford-McMahon (GM) refrigerators, which have two cold heads, were installed on the top of the cryostat. The cooling power of the GM cryocooler is 1.5 W at the second stage and 50 W at the first stage. Each stage was connected to the liquid helium reservoir, a radiation shield including high-Tc current lead, and related items. Before commissioning the ECR ion source, a preliminary evaluation of the recondensation performance was carried out with the magnet in partial operation. The design of the cryostat, its fabrication, and the experimental results are reported.

