Accuracy of water displacement hand volumetry using an ethanol and water mixture
Alan R Hargens1, Jong-Moon Kim2, Peihong Cao2
1Department of Orthopaedic Surgery, University of California, San Diego, San Diego, CA 92103-8894, USA. ahargens@ucsd.edu
Background:
The traditional water displacement method for measuring limb volume is improved by adding ethanol to water.
Methods:
Four solutions were tested (pure water, 0.5% ethanol, 3% ethanol, and 6% ethanol) to determine the most accurate method when measuring the volume of a known object.
Results:
The 3% and 6% ethanol solutions significantly reduced (P < 0.001) the mean standard deviation of 10 measurements of a known sphere (390.1 +/- 0.25 mi) from 2.27 ml with pure water to 0.9 ml using the 3% alcohol solution and to 0.6 using 6% ethanol solution (the mean coefficients of variation were reduced from 0.59% for water to 0.22% for 3% ethanol and 0.16% for 6% ethanol). The spheres' volume measured with pure water, 0.5% ethanol solution, 3% ethanol solution, and 6% ethanol solution was 383.2 +/- 2.27 ml, 384.4 +/- 1.9 ml, 389.4 +/- 0.9 ml, and 390.2 +/- 0.6 ml, respectively. Using the 3% and 6% ethanol solutions to measure hand volume blindly in 10 volunteers significantly reduced the mean coefficient of variation for hand volumetry from 0.91% for water to 0.52% for the 3% ethanol solution (P < 0.05) and to 0.46% for the 6% ethanol solution (P < 0.05). The mean standard deviation from all 10 subjects decreased from 4.2 ml for water to 2.3 ml for 3% ethanol solution and 2.1 ml for the 6% solution.
Discussion:
These findings document that the accuracy and reproducibility of hand volume measurements are improved by small additions of ethanol, most likely by reducing surface tension of water.
More Related Videos
Related Concept Videos
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Uncertainty in Measurement: Reading Instruments
Volatilization
Design Example: Application of Archimedes' Principle
The volume of seawater displaced by the block is determined by first calculating the block's weight. This is done by multiplying the...
Density and Archimedes' Principle